Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands

  • ,
  • Manuel Chavez-Ruiz
    ,
  • Jorge Inolopú-Cucche
    ,
  • Jhonatan Rabanal-Sanchez
    ,
  • Lenin Rueda-Torres
    ,
  • Gloria Sanchez-Holguin
  • Universidad Privada Norbert Wiener
    ,
  • Instituto Nacional de Salud, Lima
Research Output:
Contribution to journal
Article
Peer-review

Open access

Publication Information

Output type

Research Output:
Contribution to journal
Article
Peer-review

Original language

English

Pages from-to (Number of pages)

Pages 324-330 (7 pages)

Journal (Volume, Issue Number)

Indian Journal of Clinical Biochemistry (Volume 38, Issue 3)

Publication milestones

  • Published - 07/2023

Publication status

Published - 07/2023

ISSN

0970-1915

Publication IDs

  • Scopus: 85131036334

Abstract

Peru is one of the countries with the highest lead contamination in the world. Biological monitoring has limitations due to the shortage of laboratories with validated methodologies for the measurement of blood lead, and it is necessary to use alternative methods for its measurement in high-altitude cities. We aimed to compare the blood lead levels (BLL) measured by the LeadCare II (LC) method and Graphite Furnace Atomic Absorption Spectrometry (GF-AAS). We measured the BLL of 108 children from the city of La Oroya. The mean and median BLL for GF-AAS were 10.77 ± 4.18 and 10.44 µg/dL, respectively; for the LC method, the mean was 11.71 ± 4.28 and the median was 11.60 µg/dL. We found a positive linear correlation (Rho = 0.923) between both methods. Notwithstanding, the Wilcoxon test suggests a significant difference between both methods (ρ = 0.000). In addition, the Bland–Altman analysis indicates that there is a positive bias (0.94) in the LC method, and this method tends to overestimate the BLL. Likewise, we performed a generalized linear model to evaluate the influence of age and hemoglobin on BLL. We found that age and hemoglobin had a significant influence on BLL measured by the LC method. Finally, we used two non-parametric linear regression methods (Deming and Passing-Bablok regression) to compare the LC method with the GF-AAS. We found that these methods differ by at least a constant amount, and there would be a proportional difference between both. Although in general there is a positive linear correlation, the results of both methods differ significantly. Therefore, its use in cities located at high altitudes (higher than 2440 m.a.s.l.) would not be recommended.