TY - JOUR
T1 - Accuracy of Four Apex Locators in Premolars with Root Resorption
T2 - An In vitro Study
AU - Astrid, López Maekawa
AU - Carmen Rosa, García Rupay
N1 - Publisher Copyright:
© 2022, Iranian Association of Endodontics. All rights reserved.
PY - 2022/12/1
Y1 - 2022/12/1
N2 - Introduction: The aim of the present in vitro study was to compare the accuracy of the working length measurements of four foramen locators: Root ZX mini, Raypex 6, Woodpex III and Propex Pixi in uniradicular premolars with simulated root resorption. Materials and Methods: For this study, 30 single-rooted permanent premolars were selected. The samples were divided into two groups: 15 teeth with simulated external root resorption and 15 teeth with internal resorption. Each sample was immersed in containers with alginate covering only the root part. Measurements were acquired from the four foramen locators and compared with the working length measurements that have been previously obtained with the direct visual technique and the use of a Digital USB Microscope under 5x. Next, the Shapiro-Wilk normality test was applied. The ANOVA test and f-test for related samples were performed, in order to analyze the measurements obtained. Results: In the teeth with external resorption, no statistically significant differences in the measurements were obtained with Raypex 6 and Propex Pixi foramen locators, but there were statistically significant differences with Root ZX mini and Woodpex III Root Zx mini [Root Zx mini (P=0.040) and Woodpex III (P=0.000)]. On the other hand, in the samples with internal root resorption, there were no statistically significant differences in measurements with the Root ZX mini, Propex Pixi and Raypex 6 foramen locators. However, there were significant differences with Woodpex III. Conclusions: Based on this in vitro study, Raypex 6 had the highest accuracy in premolars with simulated external resorption and Root ZX mini was the most accurate in teeth with simulated internal resorption. Furthermore, external root resorption affects the accuracy of foramen locators more than internal resorption.
AB - Introduction: The aim of the present in vitro study was to compare the accuracy of the working length measurements of four foramen locators: Root ZX mini, Raypex 6, Woodpex III and Propex Pixi in uniradicular premolars with simulated root resorption. Materials and Methods: For this study, 30 single-rooted permanent premolars were selected. The samples were divided into two groups: 15 teeth with simulated external root resorption and 15 teeth with internal resorption. Each sample was immersed in containers with alginate covering only the root part. Measurements were acquired from the four foramen locators and compared with the working length measurements that have been previously obtained with the direct visual technique and the use of a Digital USB Microscope under 5x. Next, the Shapiro-Wilk normality test was applied. The ANOVA test and f-test for related samples were performed, in order to analyze the measurements obtained. Results: In the teeth with external resorption, no statistically significant differences in the measurements were obtained with Raypex 6 and Propex Pixi foramen locators, but there were statistically significant differences with Root ZX mini and Woodpex III Root Zx mini [Root Zx mini (P=0.040) and Woodpex III (P=0.000)]. On the other hand, in the samples with internal root resorption, there were no statistically significant differences in measurements with the Root ZX mini, Propex Pixi and Raypex 6 foramen locators. However, there were significant differences with Woodpex III. Conclusions: Based on this in vitro study, Raypex 6 had the highest accuracy in premolars with simulated external resorption and Root ZX mini was the most accurate in teeth with simulated internal resorption. Furthermore, external root resorption affects the accuracy of foramen locators more than internal resorption.
KW - Apical Foramen
KW - Electronic Apex Locator
KW - Root Resorption
UR - http://www.scopus.com/inward/record.url?scp=85128938155&partnerID=8YFLogxK
U2 - 10.22037/iej.v17i1.35498
DO - 10.22037/iej.v17i1.35498
M3 - Artículo
AN - SCOPUS:85128938155
SN - 1735-7497
VL - 17
SP - 20
EP - 26
JO - Iranian Endodontic Journal
JF - Iranian Endodontic Journal
IS - 1
ER -